Values from the Spectrally Raman Analyzers catalog. Inline means the probe sits in the process itself and reads without a sample being drawn. The configuration for Orange, wavelength, spectral range and measurement points, is set in the feasibility report.

Spectrally X1 INLINE measures composition in the installation, continuously through the run. Two channels come as standard, enough for an at-line grade check and a process stream from one cabinet.
| Measurement principle | Raman spectroscopy |
| Laser wavelength | 785 nm or 1064 nm (for strongly fluorescing media) |
| Spectral range | 300–1650 cm⁻¹ or 300–3500 cm⁻¹ |
| Spectral resolution | 8 cm⁻¹ |
| Acquisition time | 5–300 s |
| Detector | Back-thinned CCD, TEC-cooled |
| Measurement channels | Up to 2 as standard, more on request |
| Industrial protocol | Modbus |
| Calibration | Automatic, against a reference integrated in the probe |
| Laser lifetime | Min. 2 years |
| Operating conditions | 0–45 °C; 20–70% RH, non-condensing; warm-up 30 min |
| Power | 230 V AC, 50/60 Hz, grounded; 300 W |

An immersion probe for liquids, suspensions and slurries in vessels and lines, in batch or continuous operation, with a dedicated configuration for powders and granulates. The retractable self-cleaning module keeps the window clear while the analyzer stays in service.
| Probe type | Immersion, fiber-optic connection to the analyzer |
| Process pressure | Up to 16 bar |
| Process temperature | −40 to +125 °C |
| Process pH | 2–10 continuous; 1–14 short-term |
| Ingress protection | IP67 |
| Wetted materials | 316L stainless steel, fused-silica window, chemically resistant epoxy |
| Process connections | Flange DN32/40/50 PN16; ANSI 1¼″/1½″/2″ Class 150; NPT male 1¼″; Tri-Clamp 1½″ and 2″; custom design |
| Cleaning | Retractable self-cleaning module: automatic retract, rinse and return cycle |
Spectrally OS turns spectra into identity, composition and monomer values, checks each spectrum against the grade specification and keeps the record on the analyzer.
| Predictive models | Hybrid of neural networks (CNN) and classical chemometrics (PLS, PCA) |
| Spectral database | Proprietary and public, approx. 28,000 spectra |
| Operator interface | SpectrallyUI, touchscreen |
| Data export | CSV / PDF / RAW |
| User management | Role-based access control (RBAC) |
| Data storage | Local database; measurement and logging continue without network |
| Industrial protocol | Modbus to PLC, DCS, MES, SCADA or LIMS |
PASS/FAIL classification, anomaly detection and alarms against your grade specification, sent to the PLC or DCS.
Calibration, validation and maintenance: models start in the feasibility study, are retrained on your production data and re-calibrated when a recipe changes. Every change is logged.
Models move between Spectrally X1 LAB, X1 PORTABLE and X1 INLINE, so the model built during feasibility is the starting point for the line model, which is then retrained on your production data.
Four points that matter for a fluoropolymer, and two that stay with the lab.
Melt-flow index and mechanical tests describe the melt and the part, and stay on your lab methods.
Carbon-filled powders such as 7518 absorb the laser; the study sets which grades the at-line check covers.
GC and your polymer tests remain the analytical reference methods. The accuracy for Halar® grades comes from the feasibility study on your samples, measured against your lab methods.
Hope you find it interesting. Please contact us and talk about our analysis in detail.
a.lukasiewicz@gekkophotonics.com+1 804 593 0282Używamy plików cookie, aby strona działała poprawnie, mierzyć ruch i personalizować marketing tylko za Twoją zgodą.