Gekko Photonics / Spectrally X1 / Probe insertion sequence
How it works

Service the probe without depressurizing the line.

A full-port ball valve and a retraction chamber let the optical window be cleaned while the pipeline stays under pressure. Eight steps, from insertion to re-insertion.

Fig. 1 · DWG SPX1-INS-01 · Rev. A 8 steps · runs automatically, or step through it yourself
Ball valve retractable probe assembly operating cycle BLOWN CLEAN BETWEEN BATCHES FLOW → RINSE IN RINSE OUT Cleaning the optical window in line PROBE Cleaning optical window PROBE GLAND mounts & seals probe status: loose RETRACTION CHAMBER internal seals and rinse ports FULL-PORT BALL VALVE isolates the process WELD BOSS SPECTRALLY X1 — PROBE INSERTION / RETRACTION SEQUENCE GEKKO PHOTONICS — SPECTRALLY X1 INLINE FULL-PORT BALL VALVE RETRACTABLE ASSEMBLY — ILLUSTRATIVE, NOT TO SCALE SCALE: NTS SHEET: 1 OF 1 REV: A DRAWN: GEKKO CHK: — DWG: SPX1-INS-01
1. Insertion
Valve open — probe lowered through the full-port opening into the pipeline.

Gekko Photonics — Spectrally X1 INLINE — illustrative schematic

Full-port ball valve

Rotated 90 degrees it isolates the process, so the window can be cleaned in the retraction chamber while the line stays under pressure.

Retraction chamber

Holds the probe above the closed valve, with internal seals around the shaft and rinse ports for in-line cleaning.

Gland

Tightened around the probe at measuring depth to seal against process pressure, loosened before the probe moves.

Window cleaning

Rinse medium flushes the retraction chamber and cleans the window in line, with the probe in place and the process isolated.