Gekko Photonics / Spectrally X1 / Probe insertion sequence
How it works

Service the probe without depressurizing the line.

A full-port ball valve and a retraction chamber let the optical window be cleaned while the pipeline stays under pressure. Eight steps, from insertion to re-insertion.

Fig. 1 · DWG SPX1-INS-01 · Rev. A 8 steps · runs automatically, or step through it yourself
Ball valve retractable probe assembly operating cycle BLOWN CLEAN BETWEEN BATCHES FLOW → PROBE Cleaning optical window PROBE GLAND mounts & seals probe status: loose RETRACTION CHAMBER has internal seals around probe FULL-PORT BALL VALVE isolates the process WELD BOSS SPECTRALLY X1 — PROBE INSERTION / RETRACTION SEQUENCE GEKKO PHOTONICS — SPECTRALLY X1 PORTABLE FULL-PORT BALL VALVE RETRACTABLE ASSEMBLY — ILLUSTRATIVE, NOT TO SCALE SCALE: NTS SHEET: 1 OF 1 REV: A DRAWN: GEKKO CHK: — DWG: SPX1-INS-01
1. Insertion
Valve open — probe lowered through the full-port opening into the pipeline.

Gekko Photonics — Spectrally X1 Portable — illustrative schematic

Full-port ball valve

Rotated 90 degrees it isolates the process, so the gland can be opened with the line still under pressure.

Retraction chamber

Holds the probe above the closed valve, with internal seals around the shaft.

Gland

Tightened around the probe at measuring depth to seal against process pressure, loosened before the probe moves.

Window cleaning

The probe comes out through the gland for manual cleaning, which often coincides with a between-batch line purge.